PROCEDUREv2.10 U P /R/R *** WOW & FLUTTER ANALYZER OPTION PERFORMANCE CHECK PROCEDURE ***/R/R Copyright (C) 1988, 1991, 1993 by Audio Precision Inc./R /R/R This procedure checks the operation of the W&F-1 wow and flutter analyzer/R using a special test signal that contains an accurately known amount of/R squarewave FM. The generator frequency is rapidly switched between two/R measured values near 3.15 kHz using a compiled utility program. The analyzer/R readings are corrected to display the equivalent calibration with sinewave/R FM with <0.3% typical error. A properly functioning analyzer should read/R 0.950% to 1.050% to be within its specified 5% accuracy./R/R If this program fails to run, you must quit and restart S1.EXE with at least/R 80K of memory reserved for DOS./R /R/R/R/R Press to continue, to abort.../E D WFTST2/R U END