PROCEDUREv2.10 U P /R ***** DCX-127 PROGRAMMABLE DC OUTPUT ADJUSTMENT PROCEDURE *****/R/R Copyright (C) 1988, 1991, 1992 Audio Precision Inc./R/R/R/R/R/R/R A PRECISION DC VOLTMETER (<0.01%) IS REQUIRED FOR CALIBRATION. PERFORM/R ANY ADJUSTMENTS BETWEEN +20C AND +25C ONLY./R /R/R/R/R/R/R/R PRESS TO CONTINUE, TO ABORT.../E L T DCVOFF/R /F2 /R/R/R/R/R/R/R CONNECT THE PRECISION DC VOLTMETER TO DC OUTPUT-1./R ADJUST R5403 FOR 0.0 mV, +\-0.1 mV.../R/R -------------------------------------------------/R | DCX-127/R |/R |/R | R5403--> O O O O/R |/R |/R |/R | O O/R -------------------------------------------------/E L T DCV1CAL/R /F2 /R/R/R/R/R/R/R ADJUST R5404 FOR A DC OUTPUT OF +10.000 V, +\-0.003V.../R/R/R -------------------------------------------------/R | DCX-127/R |/R |/R | O O O O/R | \/R | R5404/R |/R | O O/R -------------------------------------------------/E L T DCVOFF/R /F2 /R/R/R/R/R/R/R CONNECT THE PRECISION DC VOLTMETER TO DC OUTPUT-2./R ADJUST R5503 FOR 0.0 mV, +\-0.1 mV.../R/R -------------------------------------------------/R | DCX-127/R |/R |/R | O O O O/R | \/R | R5503/R |/R | O O/R -------------------------------------------------/E L T DCV2CAL/R /F2 /R/R/R/R/R/R/R ADJUST R5504 FOR A DC OUTPUT OF +10.000 V, +\-0.003 V.../R/R/R -------------------------------------------------/R | DCX-127/R |/R |/R | O O O O <--R5504/R |/R |/R |/R | O O/R -------------------------------------------------/E L T DCV1CMRR/R U P /R DISCONNECT THE PRECISION DC VOLTMETER./R/R CONNECT DC OUTPUT-1 TO THE INPUT TERMINALS OF THE DCX-127 MULTIMETER,/R ALSO CONNECT THE "-" TERMINAL OF OUTPUT-1 TO THE "+" TERMINAL OF OUTPUT-2./R/R WHEN THE SWEEPING GRAPH APPEARS ON THE NEXT SCREEN, ADJUST R8402 FOR/R THE MINIMUM CHANGE IN OUTPUT VOLTAGE VERSUS COMMON MODE VOLTAGE. THE/R OUTPUT IS NOT PERMITTED TO CHANGE BY MORE THAN 8 mV FROM -2V TO +2V./R/R/R -------------------------------------------------/R | DCX-127/R |/R |/R | O O O O/R |/R |/R |/R | R8402--> O O/R -------------------------------------------------/R/R/R PUSH TO CONTINUE.../E /A9/E L T DCV2CMRR/R U P /R/R/R CONNECT DC OUTPUT-2 TO THE INPUT TERMINALS OF THE DCX-127 MULTIMETER,/R ALSO CONNECT THE "-" TERMINAL OF OUTPUT-2 TO THE "+" TERMINAL OF OUTPUT-1./R/R WHEN THE SWEEPING GRAPH APPEARS ON THE NEXT SCREEN, ADJUST R8502 FOR/R THE MINIMUM CHANGE IN OUTPUT VOLTAGE VERSUS COMMON MODE VOLTAGE. THE/R OUTPUT IS NOT PERMITTED TO CHANGE BY MORE THAN 8 mV FROM -2V TO +2V./R/R/R -------------------------------------------------/R | DCX-127/R |/R |/R | O O O O/R |/R |/R |/R | O O <--R8502/R -------------------------------------------------/R/R/R PUSH TO CONTINUE.../E /A9/E U P /R/R/R DCX-127 DC OUTPUTS CALIBRATED./E U END